BACKGROUND: The helmet is a well-tolerated interface for noninvasive ventilation, although it is associated with poor patient-ventilator interaction. A new helmet (NH) has proven to attenuate this limitation of the standard helmet (SH) in both bench study and healthy volunteers. The authors compared a NH and a SH in intensive care unit patients receiving noninvasive ventilation for prevention of postextubation respiratory failure; both helmets were also compared with the endotracheal tube in place before extubation. METHODS: Fourteen patients underwent 30-min trials in pressure support during invasive ventilation and then with a SH and a NH in a random order. The authors measured comfort, triggering delays, rates of pressurization (airway pressure-time product [PTP] of the first 300 [PTP300-index] and 500 [PTP500-index] ms from the onset of effort, and the first 200 ms from the onset of insufflation [PTP200]), time of synchrony between effort and assistance (Timesynch/Tineu), respiratory drive and frequency, arterial blood gases (ABGs), and rate of asynchrony. RESULTS: Compared with SH, NH improved comfort (5.5 [5.0 to 6.0] vs. 8.0 [7.8 to 8.0]), respectively, P < 0.001), inspiratory trigger delay (0.31 [0.22 to 0.43] vs. 0.25 [0.18 to 0.31] s, P = 0.007), and pressurization (PTP300-index: 0.8 [0.1 to 1.8] vs. 2.7 [7.1 to 10.0]%; PTP500-index: 4.8 [2.5 to 9.9] vs. 27.3 [16.2 to 34.8]%; PTP200: 13.6 [10.1 to 19.6] vs. 30.4 [24.9 to 38.4] cm H2O/s, P < 0.01 for all comparisons) and Timesynch/Tineu (0.64 [0.48 to 0.72] vs. 0.71 [0.61 to 0.81], P = 0.007). Respiratory drive and frequency, ABGs, and rate of asynchrony were not different between helmets. Endotracheal tube outperformed both helmets with respect to all variables, except for respiratory rate, ABGs, and asynchronies. CONCLUSIONS: Compared with a SH, a NH improved comfort and patient-ventilator interaction.

New versus Conventional Helmet for Delivering Noninvasive Ventilation: A Physiologic, Crossover Randomized Study in Critically Ill Patients

LONGHINI, Federico;CAMMAROTA, Gianmaria;VASCHETTO, Rosanna;MESSINA, ANTONIO;MAGNANI, Corrado;DELLA CORTE, Francesco;
2016-01-01

Abstract

BACKGROUND: The helmet is a well-tolerated interface for noninvasive ventilation, although it is associated with poor patient-ventilator interaction. A new helmet (NH) has proven to attenuate this limitation of the standard helmet (SH) in both bench study and healthy volunteers. The authors compared a NH and a SH in intensive care unit patients receiving noninvasive ventilation for prevention of postextubation respiratory failure; both helmets were also compared with the endotracheal tube in place before extubation. METHODS: Fourteen patients underwent 30-min trials in pressure support during invasive ventilation and then with a SH and a NH in a random order. The authors measured comfort, triggering delays, rates of pressurization (airway pressure-time product [PTP] of the first 300 [PTP300-index] and 500 [PTP500-index] ms from the onset of effort, and the first 200 ms from the onset of insufflation [PTP200]), time of synchrony between effort and assistance (Timesynch/Tineu), respiratory drive and frequency, arterial blood gases (ABGs), and rate of asynchrony. RESULTS: Compared with SH, NH improved comfort (5.5 [5.0 to 6.0] vs. 8.0 [7.8 to 8.0]), respectively, P < 0.001), inspiratory trigger delay (0.31 [0.22 to 0.43] vs. 0.25 [0.18 to 0.31] s, P = 0.007), and pressurization (PTP300-index: 0.8 [0.1 to 1.8] vs. 2.7 [7.1 to 10.0]%; PTP500-index: 4.8 [2.5 to 9.9] vs. 27.3 [16.2 to 34.8]%; PTP200: 13.6 [10.1 to 19.6] vs. 30.4 [24.9 to 38.4] cm H2O/s, P < 0.01 for all comparisons) and Timesynch/Tineu (0.64 [0.48 to 0.72] vs. 0.71 [0.61 to 0.81], P = 0.007). Respiratory drive and frequency, ABGs, and rate of asynchrony were not different between helmets. Endotracheal tube outperformed both helmets with respect to all variables, except for respiratory rate, ABGs, and asynchronies. CONCLUSIONS: Compared with a SH, a NH improved comfort and patient-ventilator interaction.
File in questo prodotto:
File Dimensione Formato  
20160100_0-00024.pdf

file ad accesso aperto

Tipologia: Versione Editoriale (PDF)
Licenza: DRM non definito
Dimensione 398.54 kB
Formato Adobe PDF
398.54 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11579/70249
Citazioni
  • ???jsp.display-item.citation.pmc??? 20
  • Scopus 40
  • ???jsp.display-item.citation.isi??? 36
social impact