The electrochemical impedance spectroscopy (EIS) technique was used for the study of the electrochemical behavior of an equiatomic NiTi alloy and an implant quality AISI 316 stainless steel type ASTM F138. Experiments were carried out using four different different test solutions: phosphate buffered saline (PBS), Dulbecco minimum essential medium (MEM), MEM fetal calf serum (FCS), and MEM fetal calf serum fibroblast cell (CELL). Specimens were finished to 600-grit SiC paper and were tested in conditions that did not provoke abrupt mechanical damage of the passive film. Bode-phase spectra showed the presence of two maxima and were fitted with an equivalent circuit characterized by two parallel combinations (R, resistance; CPE, constant phase element). The R1 and CPE1 branch was assigned to the inner compact passive film and the R2 and CPE2 branch to the external porous film. The resistance of the inner film R1, roughly corresponding to the polarization resistance (Rp), which is inversely proportional to the material’s corrosion rate, increased with the immersion time and was generally greater in PBS than in other media. With the exception of FCS solution, R1 for NiTi alloy is better or similar to that of ASTM F138

In vitro corrosion study by EIS of an equiatomic NiTi alloy and an implant quality AISI 316 stainless steel

RIMONDINI, Lia;
2006-01-01

Abstract

The electrochemical impedance spectroscopy (EIS) technique was used for the study of the electrochemical behavior of an equiatomic NiTi alloy and an implant quality AISI 316 stainless steel type ASTM F138. Experiments were carried out using four different different test solutions: phosphate buffered saline (PBS), Dulbecco minimum essential medium (MEM), MEM fetal calf serum (FCS), and MEM fetal calf serum fibroblast cell (CELL). Specimens were finished to 600-grit SiC paper and were tested in conditions that did not provoke abrupt mechanical damage of the passive film. Bode-phase spectra showed the presence of two maxima and were fitted with an equivalent circuit characterized by two parallel combinations (R, resistance; CPE, constant phase element). The R1 and CPE1 branch was assigned to the inner compact passive film and the R2 and CPE2 branch to the external porous film. The resistance of the inner film R1, roughly corresponding to the polarization resistance (Rp), which is inversely proportional to the material’s corrosion rate, increased with the immersion time and was generally greater in PBS than in other media. With the exception of FCS solution, R1 for NiTi alloy is better or similar to that of ASTM F138
File in questo prodotto:
File Dimensione Formato  
in vitro corrosion.pdf

file disponibile solo agli amministratori

Tipologia: Altro materiale allegato
Licenza: DRM non definito
Dimensione 213.09 kB
Formato Adobe PDF
213.09 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11579/22474
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 34
  • ???jsp.display-item.citation.isi??? 29
social impact